Submitting RNA
Sample Requirements
General Requirements:
- Include a DNase treatment with the RNA isolation method - this ensures sample purity and accurate quantitation
- Samples must be stored in nuclease-free ultrapure water or low TE buffer (10 mM Tris-HCl, pH 7.0-8.0, 0.1 mM EDTA) at -80°C
- RNA samples must not contain > 1 mM EDTA and must be free of salts and organic contaminants, such as phenol and ethanol since these substances can interfere with library preparation reactions
- Provide spectrophotometric absorbances from a NanoDrop or similar. A260/A280 and A260/A230 absorbance ratios indicate RNA purity. Values outside these ranges indicate the presence of contaminants
- A260/A280 ratio of >2.0
- A260/A230 ratio of 2.0-2.2+
See Evaluating Nucleic Acid Quality for NGS for more information
Illumina Stranded Total RNA Prep with Ribo-Zero Plus, Ligation:
- 1 - 1000 ng high-quality purified total RNA (RIN ≥ 9) is required in a final volume ≤ 10 μL
- 10 - 100 ng low-quality RNA (RIN ≥ 2) or FFPE (DV200 > 55%) samples
- 10 ng minimum for optimal quality & FFPE samples
- Failure to treat with DNase or inefficient DNase treatment can result in a significant fraction of intergenic reads in the sequence data
Illumina Stranded mRNA Prep, Ligation:
- 25 - 1000 ng high-quality purified total RNA (RIN ≥ 9) is required in a final volume ≤ 25 μL
Quantification
- Proper quality control and quantitation requires at least 4 μL of each sample in addition to that required for library preparation
- Each RNA sample is first evaluated on the Agilent TapeStation 4150 with either RNA or HSRNA ScreenTapes
- If the samples are of good quality and quantity, the samples will be quantified by Qubit and diluted appropriately for library preparation
- If the samples are of poor quality or quantity, the customer will be notified with recommendations before moving forward with library preparation
Submitting DNA
Sample Requirements
General Requirements:
- Samples must be stored in nuclease-free ultrapure water or low TE buffer (10 mM Tris-HCl, pH 8.0-8.5, 0.1 mM EDTA) at RT, +4C, or -20°C
- DNA samples must not contain > 1 mM EDTA and must be free of salts and organic contaminants, such as phenol and ethanol since these substances can interfere with the tagmentation reaction and result in assay failure
- Provide spectrophotometric absorbances from a NanoDrop or similar. A260/A280 and A260/A230 absorbance ratios indicate DNA purity. Values outside these ranges indicate the presence of contaminants
- A260/A280 ratio of 1.8-2.0
- A260/A230 ratio of 2.0-2.2+
See Evaluating Nucleic Acid Quality for NGS for more information
Illumina DNA Prep, Tagmentation:
- 1 - 500 ng is required in a final volume ≤ 10 μL
- 100 - 500 ng minimum for human DNA samples and other large complex genomes
- 1 ng minimum for small genomes (e.g. microbial)
- PCR amplicons must be > 150 bp - tagmentation cannot add an adapter directly to the distal end of a fragment; to ensure sufficient coverage of the amplicon target region, design primers to extend beyond the target region by 50 bp per end
Illumina DNA Prep PCR-free, Tagmentation:
- 25 - 2000 ng high-quality purified DNA is required in a final volume ≤ 25 μL
NEBNext Ultra II FS DNA, with Fragmentation:
- 0.1 - 500 ng is required in a final volume ≤ 26 μL
- DNA should be in 1X TE (10 mM Tris pH 8.0, 1 mM EDTA), however, 10 mM Tris pH 7.5–8, low EDTA TE or nuclease-free water are also acceptable
NEBNext Ultra II DNA, without Fragmentation:
- 0.5 - 1000 ng is required in a final volume ≤ 50 μL
- DNA should be sheared in 1X TE and diluted with with 10 mM Tris-HCl, pH 8.0, 1X TE, or 0.1X TE
Quantification
- Quantification by spectrophotometry is only useful as a ballpark estimate for input into a fluorometric assay
Submitting Individual Libraries
Sample Requirements
- All libraries to be quantified and sequenced must have Illumina-compatible p5 and p7 sequences and must use the standard NextSeq 1000 read and index primers
- All libraries to be sequenced in a single run can be pooled if they are of the same preparation type and each library sample has unique and compatible indexes
- The library prep method and indexes of each library must be provided
Quantification
- Each library is first evaluated and quantified on the Agilent TapeStation 4150 with either D1000 or HSD1000 ScreenTapes
- A dilution series is prepared for each library to obtain a working concentration suitable for qPCR
- Each library is quantified by qPCR on the Roche LightCycler 96 using an Illumina-compatible library quantification kit
- A 2 nM dilution is then made of each library using Illumina-provided resuspension buffer (RSB) and equal volumes are combined into pools for sequencing
Submitting Pools
Sample Requirements
- Each library in the pool must satisfy the aforementioned requirements for submitting individual libraries
- Pools must be at least 2 nM in concentration and a volume greater than 20 μL must be submitted
- Pools must be prepared with Illumina-provided resuspension buffer (RSB)
- Each library sample contained in the pool must have unique and Illumina-compatible indexes, preferably unique dual indexes (UDIs)
- The library prep method (including which commercial kit or whether custom primer sequences were used) and indexes of each library must be provided
Quantification
- Pools can be quantified by TapeStation and qPCR before loading on the sequencer
- If the customer declines and provides their own quantification, the Core is not responsible for poor run quality nor resulting differences in sample representation
Submitting Single-Cell/Nucleus Suspensions for 10X
Please see Library Preps - Single Cell for more information about sample preparation and submission requirements.